2. In the drop-down menu, select the
3. In the
dialog window, press the
system
line.
administrator PIN
The numerical keypad appears.
4. Enter the previous administrator PIN (factory setting: 00000).
5. Press the Enter key on the numerical keypad.
The entry is accepted.
6. Enter the new administrator PIN.
7. Press the Enter key on the numerical keypad.
The entry is accepted.
8. Enter the new administrator PIN again.
9. Press the Enter key on the numerical keypad.
The entry is accepted.
The message
Administrator PIN changed
10. Switch the device off and back on again.
The new administrator PIN is active.
Calibrating touchscreen display
If the device has a delayed reaction or does not react at all to the controls in
the touchscreen display being pressed, you can calibrate the touchscreen
display. To do so, proceed as outlined below.
1. Call up the Administrator menu.
2. In the drop-down menu, select the
3. In the
dialog window, press the
system
line.
touchscreen display
element.
system
button in the
Change
appears.
element.
system
button in the
Calibrate
Configuring the device • 45
English

Advertisement

Table of Contents
loading