Preventing electric shock
Preventing injuries and infections
CAUTION!
Hazard to patient, malfunction
►
Keep other medical devices, e.g. high-frequency surgical devices,
at a minimum distance of approx. 1 meter to prevent incorrect
measurement or wireless transmission interference.
►
Keep high-frequency devices such as cellphones at a minimum
distance of approx. 1 meter to prevent incorrect measurement or
wireless transmission interference.
►
The actual transmission output of high-frequency equipment may
require minimum distances of more than 1 meter. For details, go to
www.seca.com.
WARNING!
Electric shock
►
Set up devices which can be operated with a power pack so that
the power supply socket is within easy reach and the power supply
can be quickly disconnected.
►
Ensure that your local power supply matches the details on the
power pack.
►
Never touch the power pack with wet hands.
►
Do not use an extension cable and multiple outlets. This also applies
to the USB connection on the touchscreen display.
►
Make sure that the power cable is not crushed and cannot be
damaged by sharp edges.
►
Do not operate the device above an altitude of 3000 m.
WARNING!
Hazard to patient
►
Subject the device to a hygiene treatment after each measurement
(see "Troubleshooting" on page 51).
►
Ensure that the patient does not have any contagious diseases.
►
Ensure that the patient does not have any open wounds on the
palms of their hands or the soles of their feet.
►
Ensure that the device is steady and level.
►
The device is not designed as a standing aid. Assist people with
limited mobility, e.g. when they are getting up from a wheelchair.
►
Ensure that the weighing platform is dry before the patient steps
onto it.
►
Ensure that the patient has dry feet before stepping onto the
weighing platform.
►
Ensure that the patient does not step directly onto the edges of the
weighing platform.
►
Ensure that the patient steps onto the weighing platform slowly and
safely.
►
Route the network and power cables such that no one can trip over
them.
WARNING!
Risk of infection
►
Before and after every measurement, wash your hands to reduce
the risk of cross-contamination and nosocomial infections.
►
Hygienically reprocess the scale regularly as described in the re-
spective section in this document.
►
Make sure that the patient has no infectious diseases.
►
Make sure that the patient has no open wounds or infectious skin
alterations, which may come into contact with the device.
English
Safety information • 7