Menu Functions; Calling The Info Mode; Calling The Info-Off Mode; Quick Menu - Dräger X-am 5100 Technical Manual

Gas
Hide thumbs Also See for X-am 5100:
Table of Contents

Advertisement

4

Menu functions

4.1

Calling the Info Mode

In measuring mode, press the
If any warning or fault messages are present, the
corresponding information or error codes are displayed
(see "Troubleshooting" on page 19 to page 20).
Press the
key successively for the next display.
OK
The peak values and the exposure values TWA
1
STEL
are displayed.
Warning messages are displayed. Numerical codes
of warning messages: see "Warning messages" on
page 19.
key
OK
Fault messages are displayed. Numerical codes of
fault messages: see "Fault messages" on page 20.
key
OK
The peak values = the maximum measured values
are displayed.
key
OK
The average values of the exposures based on
a shift of, e.g., 8 hours (TWA) are displayed
key
OK
The short-term values (STEL) = average values of
the concentrations over the average value duration
are displayed
key
OK
The instrument is in measuring mode again
If no key is pressed for 10 seconds, the instrument reverts
automatically to measuring mode.
4.2

Calling the Info-Off Mode

When the instrument is in a deactivated state, press the
key.
The name of the gas, measuring unit and measuring
range limit value are displayed.
Pressing the
key again exits the Info Off mode (or via
OK
timeout).
1
Only when activated in the instrument configuration. Delivery status:
not activated.
Dräger X-am 5100
key for approx. 3 seconds.
OK
1
and
4.3

Quick Menu

4.3.1
Quick menu functions
Bump test, see "Performing the bump test" on page 11.
Fresh air adjustment, see "Performing a fresh air
adjustment" on page 15.
Display and deletion of the peak values, see "Quick
menu "Displaying and deleting peak values"" on
page 13.
4.3.2
Calling the Quick Menu
The fresh air adjustment and the bump test are activated in the
quick menu on delivery. The function for displaying and
deleting peak values can additionally be activated with the
Dräger CC Vision PC software.
1. In measuring mode, press the
If functions in the quick menu have been activated using
the Dräger CC-Vision PC software, you can select these
functions using the
activated in the quick menu, the instrument remains in
measuring mode.
2. You can select the activated functions of the quick menu by
pressing the
key.
Press the
key to call the selected function.
OK
Press the
key to cancel the active function and to
switch to measuring mode.
If no key is pressed for 60 seconds, the instrument
reverts automatically to measuring mode.
4.3.3
Quick menu "Displaying and deleting peak values"
After the function has been
selected, the current peak
values
are
displayed;
the special peak values
symbol appears in the
display at the same time.
1. The peak values can be
deleted by pressing the
key for 5 seconds.
OK
The next display appears.
2. Press the
key to end
OK
the function.
Menu functions
key three times.
key. If no functions have been
12
HCl
ppm
0
HCl
ppm
10133280.eps
10233280.eps
13