Menu functions
4.4

Adjustment menu

4.4.1
Functions of the adjustment menu
Fresh air adjustment, see "Performing a fresh air
adjustment" on page 15
1-button sensitivity adjustment
Standard sensitivity adjustment
4.4.2
Calling the adjustment menu
The adjustment menu can only be accessed by entering
a password.
Password on delivery: » 001 «
The default password on delivery can be changed using
the PC software Dräger CCVision.
1. In measuring mode, press the
The function for entering the password is selected.
The special symbol »
function) is displayed.
The display shows » 000 «, with the first digit flashing.
2. Use the
key to set the
flashing digit.
3. Press the
key, the
OK
second digit starts flashing.
4. Use the
key to set the
flashing digit.
5. Press the
key, the third
OK
digit starts flashing.
6. Use the
key to set the
flashing digit.
7. Press the
key to
OK
confirm the password once
it has been set completely.
8. The
adjustment
menu
functions can now be selected by pressing the
Press the
key to call the selected function.
OK
Press the
key to cancel the active function.
If no key is pressed for 10 minutes, the instrument
reverts automatically to measuring mode.
14