Description
2.4

Tests and Approvals

2.4.1
Marking
1
Serial No.
on a separate label
2.4.2
Permissible power packs
Power pack 83 22 237,
approved as Type ABT 0100
Temperature class T4
-20 °C ≤ Ta ≤ +50 °C
use with alkaline batteries
Duracell Procell MN1500
Temperature class T3
-20 °C  Ta  +40 °C
use with NiMH batteries
GP 180AAHC (1800 mAh)
use with alkaline batteries
Varta Powerone 4006
Varta Powerone 4106
Panasonic Powerline LR6
Power pack 83 18 704;
approved as HBT 0000
Temperature class T4
-20 °C ≤ Ta ≤ +50 °C
Power pack 83 22 244;
approved as HBT 0100
Temperature class T4
-20 °C ≤ Ta ≤ +50 °C
1
The year of manufacture is coded by the third capital letter of the serial number:
Y = 2007, Z = 2008, A = 2009, B = 2010, C = 2011, D = 2012, E = 2013, etc.
Example: Serial No. ARCH-0054: the third letter is C, so the year of
manufacture is 2011.
6
2.4.3
Safety instructions
WARNING
!
Read the safety measures in the Instructions for Use.
Do not replace or charge batteries in potentially
explosive areas. Danger of explosion!
To reduce the danger of explosion, do not mix new
batteries with old batteries and do not mix batteries
made by different manufacturers.
Always disconnect the instrument from the power pack
before carrying out any maintenance operations.
Substitution of components may impair intrinsic safety.
High off-scale readings may indicate an explosive
concentration.
CAUTION
!
Not tested in oxygen enriched atmospheres (>21 % O
Only use power packs ABT 0100 (Order No,
83 22 237), HBT 0000 (Order No. 83 18 704) or
HBT 0100 (Order No. 83 22 244). See marking on
power pack for approved batteries and related
temperature class.
).