MENUS WHEN CONNECTED TO Dräger Polytron Pulsar 2 TRANSMITTER

TX MAIN MENU

1 View current readings
READ
2 Enter Tx tag (and channel to match Rx)
3 Connect to new Tx or Rx
4 Alignment Mode (if no link to Rx)
The
option is blanked if data from the Receiver shows that the normal link is provided. Otherwise it toggles to
ON
The
option leads first to a menu to select the starting point for editing the tag string: the string currently in place, all characters
TAG
blank, or the string saved to memory when a tag was last sent. Often a Transmitter and Receiver are given similar tags, and those of differ-
ent Dräger Polytron Pulsar 2s within the same installation follow a numerical or alphabetic sequence. The next screen pro-
vides
and
keys to change the channel within its permitted range and the 11 characters of the tag string within the character set:
+
-
0123456789 ABCDEFGHIJKLMNOPQRSTUVWXYZ[\]./
NOTE
that the permitted range of channels is 0..7 for Transmitters labelled '4 to 60m' and 8..11 otherwise. The use of
the
is explained above.
NEW
MENUS WHEN CONNECTED TO Dräger Polytron Pulsar 2 RECEIVER

RX MAIN MENU

1 View current readings
READ
2 View current settings
SETS
3 Connect to new Tx or Rx
4 More...
more
The
option lists the Receiver's user-selected parameters, such as the measurement span and static warning current for the 4
SETS
20mA current loop output. The use of the
Interface (see below) this option becomes

RX ALIGN + ZERO

1 Alignment Mode
ALIG
2 Start Rx's self-zeroing
ZERO
3 Stop zeroing or alignment
4 More...
more
The use of the
and
ALIG
ZERO
RX MEMORY MENU
1 Enter Rx tag (and channel to match Tx)
2 Read Logger data from Rx into memory
3 Send User-Settings in memory to Rx
4 More...
more
The
option works as above.
TAG
Receiver's configuration), listing the serial numbers and tag strings of the Receivers whose data is currently held.
RX 4 TO 20mA TEST
1 Current +
+
2 Current -
-
3 Stop test
STOP
4 More...
more
You can test control equipment connected to the Dräger Polytron Pulsar 2 Receiver by imposing a current on its 4 to 20mA current loop
+
-
output. +
and -
allow the test current to be stepped up and down from 0.0 to 20.0mA in 0.5mA increments, starting from 4.0mA.
Return the Receiver to its normal output mode with
TAG
NEW
ON
NEW
option is explained above. If your connection to the Receiver is through the AI500 Digital
NEW
to break the existing link and connect to another Receiver if you wish.
LINK
STOP
is explained in "Installing and Commissioning the Dräger Polytron Pulsar 2" section of the manual.
TAG
LOG
USER
allows the choice of File1, File2 or File3 to hold the logger record (which includes the
LOG
STOP
without exiting this menu, or with
The Handheld Terminal
as appropriate.
OFF
to cycle back to 'Rx Main Menu'.
more
33