Operating procedures
4.
Operating procedures
4.1.
Power connection
Caution:
Connect the unit only to a grounded mains power socket!
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply voltage
specified on the identification plate.
Deviations of ±10 % are permissible.
• Connect the circulator with mains power cable (21a) to the mains
outlet (20).
• Connect the control cable (15) between the connectors T (15a, 15b).
• Connect the refrigerated circulator with mains power cable (21b) to
the mains socket.
4.2.
Switching on / Start - Stop
Version
F25
FP50
24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Switching on:
• Circulator and cooling machine may be turned on and off
with separate mains switches. The integrated control light will
illuminate to indicate that power has beeen applied.
i
The unit performs a self-test. All segments of the 4-digit
MULTI-DISPLAY (LED) and all indicator lights will
illuminate.
Then the software version (example: V1.00) and the
configuration (example: c225) of the unit appears.
The display „OFF" or „R OFF" indicates the unit is ready to
operate.
i
The circulator enters the operating mode activated before
switching the circulator off:
keypad control mode (manual operation, factory setting)
or
remote control mode (operation via personal computer).