Only for water cooled models:
OUT
IN
Ensure circulation of cooling water by connecting the tubing to cooling
¾"
water inlet and outlet on the rear of the refrigerated circulator.
Cooling water pressure 3,5 - 6 bar.
Cooling water temperature <20 °C.
3.2.

Bath fluids

Caution:
Carefully read the safety data sheet of the bath fluid used, particulary with
regard to the fire point!
If a bath fluid with a fire point of ≤65 °C is used, only supervised
operation is possible.
Recommended bath fluids:
Bath fluids
deionized water
mixture water/glycol
Thermal M
Thermal H
Thermal HY
Thermal H5S
Thermal H10S
Thermal H20S
Order No. 10 liters
8 940 100
8 940 102
8 940 104
8 940 106
8 940 114
8 940 108
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Temperature range
5 °C ... 80 °C
-30 °C bis 50 °C
+40 °C ... +170 °C
+20 °C ... +250 °C
-80 °C ... +55 °C
-50 °C ... +105 °C
-20 °C ... +180 °C
+0 °C ... +220 °C
Bath fluids
Thermal M
Thermal H
Thermal HY
Thermal H5S
Thermal H10S
ThermalH20S
Flash point fire point
>280 °C
>305 °C
>270 °C
>360 °C
>75 °C
>80 °C
>110 °C
>130 °C
>175 °C
>210 °C
>230 °C
>270 °C
Order No. 5 liters
8 940 101
8 940 103
8 940 105
8 940 107
8 940 115
8 940 109
MC
19