6.2.

High and low temperature warning functions

Factory settings:
High temperature limit
t High 205 °C
Low temperature limit
t Low
-55 °C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
If for a sensitive temperature application task adherence to a
working temperature value (setpoint) is to be supervised, then
set high and low temperature warning values.
In the example below, the >Setpoint< of 85 °C is surrounded
by the values > t High < 87 °C and > t Low < 83 °C. The
electronics immediately registers when the actual temperature
attains a temperature out of the limits and it follows a reaction
according to what is set in the menu item > Li < (Warning or
Alarm).
(see chapter 6.4. Change-over of the warning function to
shutdown function).
1.
Press the key
or
2.
Press enter
MULTI-DISPLAY (LED) (digit blinks).
3.
Set value:
Use the cursor keys
display until the numeral you wish to change blinks.
Use the increase/decrease arrows
selected numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
Press
to update the display immediately
until the menu item
is displayed.
to indicate the adjusted value on the
to move left or right on the
to store the selected value.
MC
to change the
31