Menu functions
7.2.
ATC Absolute Temperature Calibration, 3-point calibration
Examples:
1-point calibration
°C
T 1
T
3-point calibration
°C
T 1
T
T
1 = Original curve
T
1.
2.
36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
ATC serves to compensate a temperature difference that might
occur between circulator and a defined measuring point in the
bath tank because of physical properties.
Principle:
For ATC calibration, in steady state the bath temperature at
the location of the temperature sensor (T
at the respective adjusted working temperature. This value
is then set on the circulator in the menu >ATCalibration<
under menu item >Calwert X<.
This can be a 1-point, 2-point or 3-point calibration.
°C
M = Temperature measuring instrument with
temperature sensor (TF)
B = Bath tank with T = circulator
T
= Temperature on measuring point
°C
M
T
= Temperature on circulator
T
The menu ATC - Absolute Temperature Calibration
is structured in 2 levels.
ATC status
Calibration type: 1-, 2-, 3-point calibration
2 values per calibration point
ttx = Temperature indication on circulator
Indicates the value of the calibration point entered last.
Ctx = Calibration temperature x
Input of the new value for the calibration curve.
) is determined
M