Reset - Julabo F12-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F12-MC:
Table of Contents

Advertisement

Menu functions
Optimization instructions for the PID control parameters:
optimum setting
°C
Inappropriate settings may produce the following heat-up curves:
Xp too low
°C
Xp too high or Tv too high
°C
7.7.

Reset

Factory setting: no
42
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The heat-up curve reveals inappropriate control
settings.
t
°C
t
°C
t
Use this to reset all values to factory setting.
1.
Press the
button until the menu item
displayed.
2.
Press enter
3.
Press the keys
4.
Press enter
After the reset is carried out, the parameter is automatically
set to >no<.
Tv/Tn too low
Tv/Tn too high or Xp too high
to indicate the parameter (no).
to select the parameter (YES).
to confirm the reset.
t
t
is