Interface - Baud Rate, Handshake, Parity - Julabo F12-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F12-MC:
Table of Contents

Advertisement

Menu functions
7.5.
Interface - Baud rate, Handshake
Factory settings:
4.8 kBauds
2 = even
Hardware handshake
Example: 4.8 kbauds
Example: 2 = even
40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The display changes from
keypad control mode (manual operation) to
remote control mode (operation via personal computer).
To allow communication of the circulator with a PC or a
superordinated process control system take care the interface
parameters of both units match.
Adjustable interface parameters
Baud rate
19.2 = 19200 bauds
38.4 = 38400 bauds
Parity
0 = no Parity
1 = odd
2 = even
Handshake
0 = Xon/Xoff protocol
1 = Protocol RTS/CTS hardware handshake
Data bits = 7; Stop bits = 1
1.
Press the
button until the desired menu item is
displayed.
2.
Press enter
3.
Select the parameter with the keys
4.
Press enter
Parity
,
4.8 = 4800 bauds
9.6 = 9600 bauds
software handshake
to indicate the parameter.
to store the selected parameter.
.