Activating the Info mode

 In measuring mode, press the [OK] key for approx. 3 seconds.
– If any warning or fault messages exist, the corresponding note or error codes
are displayed (see Technical Manual).
Press the [OK] key successively for the next display. The peak values and
the exposure values TWA and STEV will be displayed.
– If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.

Info Off mode

With the instrument switched off, press the [ + ] key The name of the gas,
measuring unit, and measuring range limit value are displayed for all channels.
Pressing the [ + ] key again exits the Info Off Mode (or via timeout).

Activating the Quick Menu

 In measuring mode, press the [ + ] key three times.
– If functions in the quick menu are activated using the PC software
"Dräger CC-Vision", you can select these functions using the [ + ] key.
If no functions have been activated in the quick menu, the instrument remains
in measuring mode.
Possible functions:
1. Bump test mode
2. Fresh air calibr.
3. Display and deletion of the peak values
 Press the [OK] key to call the selected function.
 Press the [ + ] key to cancel the active function and to switch to measuring mode.
– If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.

Replacing the batteries / rechargeable batteries

Danger of explosion!
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in a hazardous area. Batteries /
rechargeable batteries are part of the Ex approval.
Only the following types may be used:
– Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline,
1)
Varta Type 4106
(power one) or
1)
Varta Type 4006
(industrial)
– Alkaline batteries – T4 – (non rechargeable!)
1)
Duracell Procell MN1500
– NiMH rechargeable batteries – T3 – (rechargeable)
1)
GP 180AAHC
(1800) max. 40 °C ambient temperature.
Charge the NiMH power pack T4 (HBT 0000 type) or T4 HC (HBT 0100 type)
using the corresponding Dräger charger. Charge NiMH single cells for battery
holder ABT 0100 in line with the manufacturer's specifications.
Ambient temperature during the charging process: 0 to +40 °C.
1) Not subject to the Metrological Performance Test BVS10 ATEX E 080X and PFG 10 G 001X.
Switch off the instrument:
 Press and hold the [OK] and [ + ] key simultaneously.
 Loosen the screw on the power pack and remove the power pack.
With battery holder (order no. 83 22 237):
 Replace alkali batteries or NiMH rechargeable batteries. Ensure correct polarity.
For the NiMH power pack T4 (HBT 0000 type) / T4 HC (HBT 0100 type):
 Completely replace the power pack.
 Insert the power pack into the instrument and tighten the screw, the
instrument switches on automatically.
WARNING
15