Data Storage Device (Dat, F-Chip) - Endress+Hauser PROline prosonic flow 93 Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for PROline prosonic flow 93:
Table of Contents

Advertisement

PROline Prosonic Flow 93
"
Endress+Hauser
6.6

Data storage device (DAT, F-Chip)

T-DAT (Transmitter-DAT)
The T-DAT is an exchangeable data storage device in which all transmitter parameters
and settings are stored.
The operator has to transfer specific parameterisation values from EEPROM to T-DAT
and vice versa ( = manual safety function). See the "Description of Device Functions"
manual for details ("T-DAT SAVE/LOAD" function, No. 1009).
F-Chip (Function-Chip)
The F-Chip is a microprocessor chip that contains additional software packages that
extend the functionality and application possibilities of the transmitter.
In the case of a later upgrade, the F-Chip can be ordered as an accessory
(see Page 99) and can simply be plugged on to the I/O board (see Page 114). After
start up, the software is immediately made available to the transmitter.
Caution!
To ensure an unambiguous assignment, the F-Chip is coded with the transmitter serial
number once it is plugged in. Thus, it can not be reused with other measuring devices.
6 Commissioning
97

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROline prosonic flow 93 and is the answer not in the manual?

Table of Contents

Save PDF