Endress+Hauser PROline prosonic flow 93 Operating Instructions Manual page 106

Ultrasonic flow measuring system
Hide thumbs Also See for PROline prosonic flow 93:
Table of Contents

Advertisement

9 Trouble-shooting
106
Type
Error message / No.
S
A / C COMPATIB.
!
# 121
No. # 2xx
Error in DAT / no data reception
S
LOAD T-DAT
!
# 205
S
SAVE T-DAT
!
# 206
S
COMMUNIC. I/O
$
# 261
No. # 3xx
System range limited exceeded
S
STACK CUR.OUT n
!
# 339...342
S
STACK FRQ.OUT n
!
# 343...346
Cause
Due to different software ver-
sions, I/O board and amplifier
board are only partially compati-
ble (possibly restricted function-
ality).
!
Note!
– The indication on the display
as notice message appears
only for 30 seconds (with listing
in "Previous system condition"
function).
– This condition can occur if only
one electronics board has
been exchanged; the extended
software functionality is not
available. The previously exist-
ing software functionality is still
working and the measurement
possible.
DAT transmitter:
Data backup (download) to
T-DAT failed or error when
accessing the values stored in
the T-DAT (upload).
No data reception between
amplifier and I/O board or faulty
internal data transfer.
The temporarily buffered flow por-
tions (measuring mode for pulsat-
ing flow) could not be cleared or
output within 60 seconds.
PROline Prosonic Flow 93
Remedy / spare part
Module with lower software ver-
sion has either to be actualizied by
FieldTool with the required soft-
ware version or the module has to
be replaced.
Spare parts
Page 114
1. Check whether the T-DAT is
correctly plugged into the
amplifier board
Page 116
2. Replace T-DAT if defective.
Spare parts
Page 114
Before replacing the DAT ,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3. Replace measuring electron-
ics boards if necessary. Spare
parts
Page 114
Check the BUS contacts
1. Change the upper or lower
limit setting, as applicable.
2. Increase or reduce flow, as
applicable.
Recommendations in the event of
fault category = FAULT MESSAGE
($):
– Configure the fault response of
the output to "ACTUAL VALUE"
(see Page 112), so that the tem-
porary buffer can be cleared.
– Clear the temporary buffer by
the measures described under
Item 1.
Endress+Hauser