Uv Transilluminator; Lamps Replacement (P/N 100-1361) - BIO RAD Gel Doc 2000ChemiDocChemiDoc XRS Hardware Instruction Manual

Table of Contents

Advertisement

This unit is protected by 2 fuses 5X20 mm 2A Slow Blow. The fuses are located in the rear left
side in a fuse holder. See above picture.
1. Unplug the main power cable from the power entry module outlet.
2. Using a screwdriver lever on the fuse holder to extract it.
3. Remove the blown fuses and replace them with two new ones (P/N 100-1951).
4. Slide the fuse holder into the power entry module until it snaps in place

8.3 UV Transilluminator

8.3.1 Lamps replacement (P/N 100-1361)

Attention: The UV filter surface should always be kept clean from the chemical agents used
as gel dyes. Please use protective gloves when touching the UV transilluminator cover.
The lamps must be replaced after every 500 hours of use.
Turn off the power.
Remove the power cable.
Remove the four screws located on the left/right sides of the drawer.
Remove the cover with the UV filter.
Place it down on its backside on a non-abrasive surface.
Caution: Do not put the UV filter directly on the bench.
Use glove when touching the lamps. Remove the lamp by rotating it until it
becomes loose and the pins come to a vertical position;
Remove the lamp. Install the new lamp by rotating so that the pins are horizontal
and the lamp is tight.
Reassemble the lid and retighten the screws on both sides.
Lamp replacement
UV Transilluminator
Starter replacement
Page 34 of 42