9. See picture below:
10. Next you will be prompted to remove your sample from the transilluminator and place the
Fluorescent Reference Plate on the platen for UV illuminated images. For White Light
Illumination the white light transilluminator or the UV conversion screen is used to collect
the reference Flat Fielding image. In either case you must remove the sample from the
platen.
NOTE: For White Light Conversion screen you must turn ON the UV Transilluminator
after removing the Sample. To achieve good focus you may need to close the aperture till
proper focus is achieved.
11. Remove the sample from the Transilluminator.
transillumination. Place the Fluorescent Reference Plate on the platen. Make sure that the
plate covers the entire UV filter glass. Close the Universal Hood door and then Click on
"Continue" in the window shown above.
12. Next prompt will require you to turn ON the UV Transilluminator.
Transilluminator/White Trans button on the universal hood as the case may be and click on
"Continue"
Clean the platen properly for UV
Page 27 of 42