Fluke 6060B Instruction Manual page 46

Synthesized rf signal generator
Hide thumbs Also See for 6060B:
Table of Contents

Advertisement

INSTALLATION
AND
OPERATION
I
condition.
Table
2-5
contains a
list
and
explanation of
all
the
Uncalibrated Error
Code
Messages. Table
2-6
contains a
list
and
explanation of
all
the
Rejected
Entry Error
Code
messages.
Table
2-5.
UNCAL
Error
Codes
CODE
DESCRIPTION
000 000 000 = Indicates no UNCAL conditions
001
000 000
s FM
deviation
<
100
Hz
<<10
Hz
with
Option
-651)
002 000 000
= FM
deviation
>
Cfreq
-100 kHz)/3
*004 000 000
=
Excess
FM
deviation, main
or
reference PLL unlocked
020 000 000
-
AM depth
>
90%
*000 010 000
=
Main
or
reference PLL unlocked
000 000
001 =
Level vernier below calibrated range
or
Level <
-127 dBm
000 000 002
=
Peak
(AM)
amplitude
>
+13
dBm
*000 000 004
=
Amplitude unleveled
*000 000 010
s
Fixed-range
level
vernier
at 0
*000 000 020
s
Fixed-range
Level
vernier
at
full scale
*000 000 040
s
RPP tripped
000 000 100
s
Level below -127 dBm
000 000 200
s
Level correction disabled
*000 000 400
RF
output off
Table
2-6.
REJect
ENTRY
Codes
CODE
DESCRIPTION
001
000 000
FM
deviation
not
between
0
and 99.9 kHz
<9.99
kHz with
Option
-651)
002 000 000
— FM
deviation Step
not
between
0
and
99.9
kHz
<9-99 kHz with Option -651)
004 000 000
-
AM depth not between
0
and 99%
010 000 000
-
AM depth step not between
0
and 99%
020 000 000
=
IEEE-488 command syntax error
040 000 000
-
IEEE-488
input
value
out of range
100 000 000
-
Special function requires IEEE option
200 000 000
IEEE edit or step
operation beyond allowed
range
000
001
000
=
Frequency
not
between
0-01 and 1050 MHz
000 004 ooo
=
Frequency step not between
0
and 1050 MHz
000 040 000
-
Invalid memory location
000 100 000
=
Invalid data
in
memory
000 200 000
Si
Special function not allowed
000 000
001 s
Output amplitude not between 10
nV and 2V
000 000 002
-
Insufficient resolution
for units
conversion
000 000 004
Units conversion to volts not allowed with reference
in
volts
000 000 010
-
Units conversion to
dB not
allowed with reference
in
volts
000 000 020
-
Amplitude step not between 0 and 166
dB or 0 and
1999V
000 000 040
-
Units conversion on amplitude step not allowed
000 000 100
s
Amplitude step and current amplitude display
not in Same units
2-25

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents