MAINTENANCE
CALIBRATION
ADJUSTMENTS
PROCEDURE
Adjust
R16
for
0.98V
at
TPl.
i.
Gain
access
to
the rear
of
the
Display
PCA
by removing
the
top instrument
2.
Connect
the
DVM
to
measure
the
dc
voltage
at
TPl
relative to
the
chassis.
3.
Adjust
R
16 for
+0,9800
±
-0005V
dc.
4C-5.
OUTPUT
ASSEMBLY,
A2A4,
ADJUSTMENT
This procedure covers
all
of
the
adjustments
on
the
A2A4
Output
PCA,
as follows:
1.
R309,
LEVEL
D AC
offset
2.
R419, modulation
oscillator
level
3.
R
1
44, linearizer
detector
offset
4.
R421,
AM
depth
5.
R31LRF
level
6.
R227, Het
level
These adjustments,
as well as
TP7,
are accessible
by removing
the
seven
number
10
access
screws
in
the
module
cover.
Refer
to
Figure
4C-2
to identify
the access
screw
corresponding
to
a
particular
adjustment.
Any
adjustment can be
made
independently
unless
it
is
noted
that
it
interacts
with
another adjustment. Interdependent adjustments
must
be
done
in
the
sequence
presented.
If
more
than
one
adjustment
is
necessary,
do them
in
the
sequence
presented.
1.
Level
DAC
Offset
Adjustment
TEST
EQUIPMENT
DVM
REMARKS
This adjustment
is
normally
required only
when U302
or
any
associated
components
are
replaced or
when
the
adjustment has been
changed
or
has
shifted.
CAUTION
This adjustment
directly affects
the output
level
and
should
not
be
adjusted
Indiscriminately.
PROCEDURE
The
LEVEL
DAC
Offset,
R309,
is
adjusted
for
0
±
0.5
mV
at
TP7
with
the
RF OUTPUT
turned
OFF.
a.
Gain
access
by removing
the
bottom
instrument cover
and removing
the
access
screws
for
TP7
and
R309.
b.
Program
the
UUT
to
[RCL]
[9][8],
and program
the
RF OUTPUT
to
OFF.
c.
Connect
the
DVM
to
measure
the voltage
between
TP7
and
the
power
distribution
connection
point
on
the
module
plate.
4C-4