Low-Level Accuracy Test - Fluke 6060B Instruction Manual

Synthesized rf signal generator
Hide thumbs Also See for 6060B:
Table of Contents

Advertisement

MAINTENANCE
PERFORMANCE
TESTS
I
H
OUTPUT POWER
ATTENUATION
PROGRAM
measured
ERROR
(dB)
SECTION
LIMIT
(dB)
SECTION
NOMINAL
(dBm)
(dBm)
(dB)
0
0
+12
+12.2
+0.2
+12.2-12.0
=
-K).2
1
6
+ 6
+05.9
-0.1
-12.2+5.9+6
=
+0.3
2
12
0
"0D.2
-0.2
-12.2-0.2+12
=
-0.4
3
24
-12
-12-1
-Q.1
"12.2-12.1+24
= -0.3
4
24
"12
CSPCLKS:
LZ2
-11.8
+0.2
-12.2-11.8+24
=
+0.0
5
24
-12 CSPCL]CS]C4] -12.0
+0.0
-12.2-12.0+24
=
-0.2
6
24
"12 tSPCL3C83
C53
-12.3
-0.3
-12.2-12.3+24
-
"0.5
7
.
24
-12 cspcl:c8:c6: -11.9
+0.1
-12-2-11-9+24
=
-0,1
Sum
of
Errors
=
-1,0
Figure 4A-3. High-Level
Accuracy
Test Conditions
4A-7.
LOW-LEVEL
ACCURACY
TEST
The Power
Meter
with a
Low-Level Sensor and
the calibrated
60-dB
Attenuator
are
used
to verify the
U
UT
level
accuracy
at -1
27
dBm
and
at
frequencies of
1
20,
244, 245, 850,
and
1050
MHz,
by
using
the
Spectrum
Analyzer
as
an
indicator.
REQUIREMENT
Amplitude
accuracy
is
±1.0
dB
from
+13
to
-127
dBm.
TEST
EQUIPMENT
Wideband
Amplifier
60-dB
Attenuator
RF
Spectrum
Analyzer
Power
Meter
with a
Low-Level Sensor
REMARKS
This
test,
in
conjunction with
the
Mid-Level Accuracy and High-Level Accuracy
Test,
verifies
the overall
level
performance of
the
UUT.
Failing
this
test,
but passing
the
High-Level Accuracy
Test,
probably
indicates
a
leak-
around problem
in
the
UUT
attenuator. Service
tip:
Check
for
a broken
feed-through
filter
or
improper mechanical
assembly,
i.e.
,
loose
screws
and/or
damaged
or misplaced
gaskets.
It
is
convenient
to
use the
UUT RF ON/OFF
control
when
zeroing the
Power
Meter.
4A-7