MAINTENANCE
PERFORMANCE
TESTS
This
test verifies
the high-level
accuracy of
the
Generator
and
also
verifies
that the
amplitude
correction factors for the individual
Attenuator
sections are correct.
This
test,
in
conjunction with
the mid-level
accuracy
and
low-level
accuracy
tests,
verifies
the
overall
level
performance
of
the
UUT.
PROCEDURE
a.
Calibrate
and
zero
the
Power
Meter.
b.
Program
the
UUT
to
[RCL]
[9][8].
c.
Connect
the
Power
Sensor
to
the
UUT RF
OUTPUT.
d.
Program
the
UUT
frequency
to 0.4
MHz.
e.
Select
each attenuator
section
by
programming
the
UUT
amplitude
to
the
levels
shown
in
Figure
4A-2
High-Level
Accuracy
test
conditions,
and
record the
measured
power
at
each
level.
f.
For
each
programmed
level
of Figure
4A-2,
compute
the
output
power
error
(subtract the
programmed power
in
dBm
from
the
measured power
in
dBm).
These
errors
must
not exceed
±
1.0
dB.
g.
For
attenuator
sections
1
through
7,
subtract the
measured power
for section
zero
from
the
sum
of the
measured power
for that section
and
the
nominal
attenuation for that
section,
e.g.,
(-M0+M1+6)
for section
1.
The
eight section
errors
and
their
sum
must
not exceed
±
1.0
dB.
Figure
4A-2 shows
the
parameters of
the high-level
accuracy
test.
OUTPUT POWER
ATTENUATIOM
program
MEASURED
ERROR
(d8)
SECTION
ERROR
<dB>
LIMIT
(dB)
SECTION
NOMINAL
CdBm)
(dBm)
0
0
+12
MO
MO-12
MO-12
±1 .0
1
6
+ 6
Ml
MI-6
-M0+M1+6
2
12
0
M2
M2-0
-MO+M2+12
M
3
24
-12
M3
M3-12
-M0+M3+24
II
4
24
-12
CSPCL1 L&2ZZ2 M4
M4-12
-M0+M4+24
H
5
24
-12
CSPCL3 ES:C4:
M5
M5-12
-MD+M5+24
II
6
24
-12
cspcl: tSlL52 M6
M6-12
-M0+M6+24
f|
7
24
-12
CSPCL] :S3:6] M7
M7-12
-MQ+M7+24
"
Sum
of
Errors
±1-0
Figure 4A-2. High-Level
Accuracy
Test Conditions
4A-5