INSTALLATION
AND
OPERATION
part includes typical
timing
data,
provided
as
an
aid to
system
programmers.
This
information can
assist in
writing
programs
that
have
greater
speed
and
efficiency.
The
Generator can
be
used with
any IEEE-488
controller
in
the
normal
addressed
mode.
The
following
two
additional
modes
are available for
operation without a
controller:
Listen-only
mode
Talk-only
mode
In
the listen-only
mode,
the
Generator responds
to
all
data messages
on
the
IEEE-488
bus. In the talk-only
mode,
the
Generator
sends
commands
on
the
IEEE-488
bus
to
program
another
6060B
Generator
(or a
607xA
with
some
restrictions).
2-34.
Setting
Up
the
IEEE-488
Interface
Figure
2-5
shows
a
6060B
Signal
Generator connected
to a
1
722A
via the
1
EEE-488
bus.
Rgure
2-5.
6060B
Signal
Generator
Connected
to
a
1
722A
Use
the
following
procedure
to
set
up
the
Generator
with
the
IEEE-488
Interface option:
1.
Connect
a
standard
IEEE-488
cable
between
the
Generator
and
the
IEEE-488
device.
NOTE
The IEEE-488
Interface signal
SHIELD
(pin 12)
can he disconnected
(when
using
an IEEE-488
cable with
a
tneiallic
hood)
from
instrument ground.
To do
this,
use the
left
most
address switch
(as
viewed
from
the rear panel).
2.
Select the
IEEE-488
address
and
mode
as follows:
a.
For
control of the
Generator
with
a
controller,
set
both
the
LISTEN
ONLY
and
TALK ONLY
switches
to
0 (down). Set switches
al
through
a5
to the
desired
address 0
through
30.
For
example,
for
an address
of
I
,
set
switches
a2,
a3
a4,
and
a5
to
0 (down),
and
set
switch
al to
1
(up).
b.
For
talk-only operation,
set
the
TALK ONLY
switch
to
1
(up).
c.
For
listen-only
operation,
set
the
LISTEN
ONLY
switch
to
1
(up)
and
the
TALK ONLY
switch
to
0 (down).
3.
Verify the address
and mode:
2-28