Fluke 6060B Instruction Manual page 18

Synthesized rf signal generator
Hide thumbs Also See for 6060B:
Table of Contents

Advertisement

INTRODUCTION
AND
SPECIFICATIONS
Table
1-1.
Signal
Generator
Specifications (cont)
POWER
100. 120, 220. 240V
ac
+10%,
47
to
63
Hz,
<180 VA
(<15 VA,
with Option
"130 instaLled,
and the
Generator
turned off (standby).
WEIGHT
<15.7
kg (35 Lbs).
NON-VOLATILE MEMORY
50 instrument states
are retained
for
2
years (typically), even without
ac
Line
power applied.
reverse power protection
PROTECTION LEVEL
Up to 50 watts from
a
50-ohm
source,
0.01 MHz to
1050
MHz.
Withstands
up
to
50V
dc.
Protection
is not
provided
when the Generator
is
off.
606OB OPTION -130
HIGH
STABILITY REFERENCE
AGING RATE
<±5
x
10"'*°/day,
after
21
days
continuous operation.
TEMPERATURE STABILITY
<±2
x
(Oven remains
powered
in
standby).
6Q60B OPTION -132 MEDIUM STABILITY REFERENCE
AGING RATE
<±1
x
10"^/month after
5
days
continuous operation.
TEMPERATURE STABILITY ...........
<+1
x
1Q"^/°C
(0
to 50®C)
(no
powered standby).
6060B OPTION -48S IEEE-488 INTERFACE
(IEEE Std 4S8-197S)
INTERFACE FUNCTIONS
SHI,
AH1
,
T5, TEO, L3, LEO, SRI,
RL1
,
PPO, DC1, DTI,
CO,
and
El.
6060B OPTION
"651 LOW RATE
EXTERNAL
FM
MAXIMUM DEVIATION
DROOP
BANDWIDTH
(3dB)
MAX DC INPUT
INCIDENTAL
AM
9.99
kHz.
<15% on
a
10-Hz square wave.
0-5
Hz
to
100
kHz
(typical).
±10
mV.
<1% AM
at
1-kHz rate and
deviation <10
kHz.
SUPPLEMENTAL CHARACTERISTICS
The
following characteristics are provided
to assist in the
application
of the
Generator
and
to describe the typical performance that
can be
expected.
FREQUENCY SWITCHING SPEED
<100
ms to be
within 100
Hz.
1-8