SECTION 11 – TROUBLESHOOTING
11.1

CALIBRATION ERROR CODES

If an error code is displayed during calibration it will be accompanied by a Clipboard icon and a
number to indicate the cause of the error. The first time that an error is displayed, it will be possible
to repeat the erroneous measurement. If the same error message is shown after the repeat
measurement, the complete calibration process will need to be repeated. The table below shows
the error codes:

11.1.1 Calibration Procedure Error Codes

Error Code
Symbol
1
3
5
7
8

11.1.2 Verification Check Error Codes

9
11
Issue
Blank solution reading outside of acceptable range
Calibration solution reading outside of acceptable range
Duplicate calibration solution readings outside of acceptable
range
0.2mm path length factor outside of acceptable range
0.5mm path length factor outside of acceptable range
Blank solution reading outside of acceptable range
Calibration solution reading outside of acceptable range