Section 7 - Calibration Of The Micro Volume Accessory; Calibration Solutions (035 092); Micro Volume Calibration Procedure; Micro Volume Settings - jenway Genova Nano Operating Manual

Table of Contents

Advertisement

SECTION 7 – CALIBRATION OF THE MICRO VOLUME ACCESSORY
Jenway recommends that the micro volume accessory is calibrated every 6 months. A set of calibration
solutions (Part code 035 092) are supplied with the Genova Nano spectrophotometer. Please note that
the calibration solutions should be discarded 1 week after being opened.
When using the calibration solutions we advise the use of chemically resistant gloves and googles and
that there is eye wash available immediately. We recommend safe handling of the calibration solution -
avoid skin contact, direct inhalation or ingestion of the standards as advised in the MSDS.
7.1

CALIBRATION SOLUTIONS (035 092)

7.2

MICRO VOLUME CALIBRATION PROCEDURE

7.2.1

Micro Volume Settings

7.2.2

Calibration Standard - Data Entry

The supplied calibration solution set consists of two
vials.
1. Matrix Blank (White)
2. 10x ref - Calibration Standard (Blue)
A calibration certificate is supplied that details the
certified absorbance values of the calibration
standard and the path length at which these values
were determined.
The micro volume settings can be accessed by
pressing the key below the Micro Volume icon in
any of the Genova Nano's operating modes.
1. Enter the A260 and A330 DNA standard
solution values given on the calibration
certificate into the accessory settings
pressing the key adjacent to the Calibration
Standard Value icon.
by
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents