Manual Setup - How to change scale of Scope Display
The example below shows stepwise how to adjust Scope display scaling of the phase
voltages.
info@Fluke-Direct.ca
Use the arrow keys to highlight 3φ IT. Then press
ENTER to confirm the selection.
ENTER
The diagram shows in detail how to connect
voltage and current probes to the power system
under test.
F5
(3x)
Return to Setup entry screen. The new
configuration is indicated behind Config. and the
belonging configuration symbol is shown on the
right side of the screen.
The voltage waveforms of the Phases are outside
the viewing window.
Fluke-Direct
Setting up the Analyzer
MANUAL SETUP
.ca
1.800.868.7495
24
24-11