Operation
20
4.2.5 AutoRead AR (drift control) and hold function
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
Use the AutoRead function and hold function like this:
1
Select the required measuring mode with <M>.
2
Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3
Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
pH
S
5 3
C
O
Sal
Time
Baud
0
2 5 0
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
4
If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5
To terminate AutoRead: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.