Tds (Total Dissolved Solids) - wtw Cond 1970i Operating Manual

Portable conductivity measuring instrument
Table of Contents

Advertisement

Cond 1970i
ba75692e02
07/2009

4.2.4 TDS (Total dissolved solids)

You can measure the total dissolved solids as follows:
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the conductivity measuring cell in the test sample.
3
Press the <M> key until the unit mg/l appears.
The value of the total dissolved solids appears in the upper dis-
play line. The TDS factor appears in the lower display line.
pH
S
C
4 7
O
Sal
Time
Baud
0
1 0 0
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
4
Using <▲> <▼>, set the TDS factor (0.40 ... 1.00).
(The TDS factor has to be determined by a comparative mea-
surement before.)
pH
S
3 9
C
O
Sal
Time
Baud
0
0 8 2
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
5
Wait for a stable measured value.
mV
/
pH
4
%
mg/l
S/
m
cm
1/
Sal
cm
C
°
/
%
K
TP
REL1
REL2
ARng
AR
RCL
mV
/
pH
8
%
mg/l
S/
m
cm
1/
cm
Sal
C
°
/
K
%
TP
REL1
REL2
ARng
AR
RCL
Operation
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents