Pc/External Printer (Rs232 Interface); Remote Control - wtw Cond 1970i Operating Manual

Portable conductivity measuring instrument
Table of Contents

Advertisement

Cond 1970i
Socket assignment
ba75692e02
07/2009

4.6.3 PC/external printer (RS232 interface)

Via the RS 232 interface, you can transmit the data to a PC or an ex-
ternal printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the
interface to the devices.
The data output automatically switches to the RS232 interface.
Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
Set up the following transmission data at the PC/printer:
Baud rate
selectable between: 1200, 2400, 4800, 9600
Handshake
RTS/CTS + Xon/Xoff
Parity
none
Data bits
8
Stop bits
1
1 CTS
2 RxD
1
2
3 Ground
4
3
4 TxD
RS 232
REC

4.6.4 Remote control

The measuring instrument can be remotely controlled from a PC. This
requires the KOM pilot communication kit. It is available as an acces-
sory.
The instrument is then controlled via commands that simulate key-
strokes and request the current display contents.
Note
A more detailed description is provided within the scope of delivery of
the communication kit.
Operation
45