Determining The Cell Constant (Calibrating) - wtw Cond 1970i Operating Manual

Portable conductivity measuring instrument
Table of Contents

Advertisement

Cond 1970i
ba75692e02
07/2009

4.3.1 Determining the cell constant (calibrating)

Note
This method of automatically determining the cell constant by calibra-
tion with the 0.01 mol/l KCL standard solution can only be used for
measuring cells with cell constants in the range 0.450 ... 0.500 cm
-1
0.800 ... 1.200 cm
.
This is how you can determine the cell constant:
1
Press the <CAL> key until CELL appears on the display.
pH
S
1
C T 1
O
Sal
Time
Baud
C
E L L
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
2
Press the <RUN/ENTER> key.
3
Press the <CAL> repeatedly, until the calibrated cell constant
appears on the display: CAL appears on the display.
pH
S
C
C
A
O
Sal
Time
Baud
0
4 7 5
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
4
The displayed value is the current calibrated cell constant. You
can:
– accept this setting for measuring with <M> or
– continue with step 5 and start a new calibration.
5
Immerse the measuring cell in the control standard solution,
0.01 mol/KCl.
mV
/
pH
%
mg/l
S/
m
cm
1/
cm
Sal
C
°
/
K
%
TP
REL1
REL2
ARng
AR
RCL
mV
/
pH
L
%
mg/l
S/
m
cm
1/
Sal
cm
C
°
/
K
%
TP
REL1
REL2
ARng
AR
RCL
Operation
-1
or
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents