4-20
Range, Digits, Rate, Bandwidth, and Filter
Remote programming — filter
Filter commands
The filter commands are listed in
follow the table.
NOTE
Table 4-6
Filter commands
Commands
DCV filter commands
[SENSe[1]]
:VOLTage[:DC]:AVERage:TCONtrol <name>
:VOLTage[:DC]:AVERage:WINDow <NRf>
:VOLTage[:DC]:AVERage:COUNt <n> [, clist]
:VOLTage[:DC]:AVERage:STATe <b> [, clist]
ACV filter commands
[SENSe[1]]
:VOLTage:AC:AVERage:TCONtrol <name>
:VOLTage:AC:AVERage:WINDow <NRf>
:VOLTage:AC:AVERage:COUNt <n> [, clist]
:VOLTage:AC:AVERage:STATe <b> [, clist]
DCI filter commands
[SENSe[1]]
:CURRent[:DC]:AVERage:TCONtrol <name>
:CURRent[:DC]:AVERage:WINDow <NRf>
:CURRent[:DC]:AVERage:COUNt <n>
[, clist]
:CURRent[:DC]:AVERage:STATe <b> [, clist]
ACI filter commands
[SENSe[1]]
:CURRent:AC:AVERage:TCONtrol <name>
:CURRent:AC:AVERage:WINDow <NRf>
:CURRent:AC:AVERage:COUNt <n> [, clist]
:CURRent:AC:AVERage:STATe <b> [, clist]
Query commands are not included in
subsystem are provided in
1, 4
Table
4-6. Additional information on these commands
Table
4-6. All commands for the SENSe
Table
15-5.
Description
Optional root command.
Select filter type; <name> = MOVing or
REPeat.
Set filter window in %; <NRf> = 0 to 10.
Specify filter count; <n> = 1 to 100.
Enable or disable the filter.
Optional root command.
Select filter type; <name> = MOVing or
REPeat.
Set filter window in %; <NRf> = 0 to 10.
Specify filter count; <n> = 1 to 100.
Enable or disable the filter.
Optional root command.
Select filter type; <name> = MOVing or
REPeat.
Set filter window in %; <NRf> = 0 to 10.
Specify filter count; <n> = 1 to 100.
Enable or disable the filter.
Optional root command.
Select filter type; <name> = MOVing or
REPeat.
Set filter window in %; <NRf> = 0 to 10.
Specify filter count; <n> = 1 to 100.
Enable or disable the filter.
Model 2701 User's Manual
5
Default
(Note 2)
0.1
10
(Note 3)
(Note 2)
0.1
10
(Note 3)
(Note 2)
0.1
10
(Note 3)
(Note 2)
0.1
10
(Note 3)