Amps Measurement Procedure - Keithley 2701 User Manual

Ethernet-based dmm/data acquisition system
Hide thumbs Also See for 2701:
Table of Contents

Advertisement

3-18
Basic DMM Operation
Model 7700 switching module
Connections for the Model 7700 switching module are shown in
only channels 21 and 22 can be used for current measurements.
Figure 3-7
DCI and ACI connections using Model 7700 switching module
Current
Source
Caution: Maximum input: 60VDC or 30V RMS, 3A switched,

Amps measurement procedure

NOTE
1.
2.
3.
4.
CAUTION
5.
CH 21 or 22
60W, 125VA maximum.
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it has to be in the "F" (out) position. For switching modules, it has to be
in the "R" (in) position.
If a switching channel is presently closed (displayed), press OPEN to open it.
Select the amps measurement function by pressing DCI or ACI.
Use the RANGE
and
expected current or press AUTO to select autoranging (AUTO annunciator turns
on). Details on range are provided in
Apply the current(s) to be measured.
Do not apply more than 3A to the input or the AMPS fuse will blow.
Model 7700 switching module — When performing current measure-
ments, the maximum allowable voltage is 60VDC or 30V RMS.
Exceeding these limits could cause damage to the switching module.
If using a switching module, use the
channel (for the Model 7700, 21 or 22). All other channels will be open.
H
Model 7700
Switching
Module
L
keys to select a measurement range consistent with the
Section
4.
and
keys to close the desired amps
Model 2701 User's Manual
Figure
3-7. Note that

Advertisement

Table of Contents
loading

Table of Contents