Temperature Measurement Procedure - Keithley 2701 User Manual

Ethernet-based dmm/data acquisition system
Hide thumbs Also See for 2701:
Table of Contents

Advertisement

Model 2701 User's Manual

Temperature measurement procedure

NOTE
1.
2.
3.
4.
5.
NOTE
6.
7.
8.
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the "F" (out) position. For switching modules, it must be in
the "R" (in) position.
If a switching channel is presently closed (displayed), press OPEN to open it.
Select the temperature measurement function by pressing TEMP.
Configure the temperature measurement as previously explained in
measurement configuration," page
Connect the temperature transducer(s) to be measured.
If using a switching module, perform the following steps to close the desired
channel. Keep in mind that for 4-wire RTD measurements, you will close the
primary (INPUT) channel (1 through 10). The channel that it is paired to will close
automatically.
a.
Press the CLOSE key.
b.
Use , , , and
previously closed channel(s) (if any) will open and the specified channel (or
channel pair) will close.
While in the normal measurement state, you can use the
channels. In general, each key press will open the presently closed channel and
then close the next higher or lower channel.
Observe the displayed reading.
To measure other switching channels, repeat steps 5 and 6.
When finished, press OPEN if there is a channel closed.
3-42."
to key in the channel number and press ENTER. The
Basic DMM Operation
"Temperature
and
keys to close
3-45