Fluke 433 User Manual page 92

Three phase power quality analyzer
Hide thumbs Also See for 433:
Table of Contents

Advertisement

Fluke 433/434
Users Manual
Limits Adjustments is used to save, recall, and define sets of limits for:
Read below how to do this:
18-14
Power Quality Monitoring.
Dips/Interruptions/Rapid Voltage Changes/Swells.
Adjust Monitor Limits is the entry menu. It shows the main settings of
1
the active set of limits: name, creation date, and a summary of limits
data.
If required, use the arrow keys to define a name for a set of limits that
you want to save.
Recall Monitor Limits menu is used to recall a set of power quality
2
limits. A maximum of six sets can be recalled:
- Default 1 and 2 are factory installed read-only sets: one of them is the
set of limits according to the EN50160 standard.
- Admin 1 and 2 are sets definable by an administrator by means of PC-
software: for the user these sets are read-only.
- User 1 and 2 can be defined and saved by the user.
Use the up/down arrow keys to select a set of limits you want to recall.
Then press Function key F5 to recall and to use them.
Press Function key F1 to leave the menu without further actions.
Edit Monitor Limits menu is used to modify limits. Setups are grouped
3
per power quality item in separate submenus for voltage, harmonics,
flicker, etc.
Use the up/down arrow keys to select an item to be adjusted. Then press
the ENTER key F5 to enter the adjustment submenu. All adjustment
items are listed in the table below.
Use the arrow keys to select and edit limits.
4
Press Function key F5 to confirm selections and return to the Edit
Limits menu. Use Function keys F1 – PREVious or F2 – NEXT to
move directly to an adjacent submenu. When ready with editing the
limits, Press Function key F5 – OK twice to return to the Adjust
Monitor Limits menu. Arrow keys can be used here to define a name
for the new set of limits. Then press Function key F2 – SAVE to enter
the Save Monitor Limits menu.
Save Monitor Limits menu is used to save sets of limits in User 1 or 2.
5
Use the up/down arrow keys to select User 1 or User 2. When available
save the set of limits in an empty location; saving into a location
already filled will overwrite the existing set. Press Function key F5 –
SAVE to do the save action. Press F1 – CANCEL to return to the
Adjust Monitor Limits menu without saving limits. In this menu you
can also define a name for a set of limits to be saved.