MAXCAP DI MAINTENANCE AND REPLACEMENT:
Procedure:
1.
When the reading on the right-hand TDS meter (set to "IN") displays 50% of
the reading on the left-hand TDS meter (set to "OUT"), it is time to replace
the MAXCAP
2.
Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
3.
Tighten the cartridge housing by rotating it clockwise and hand tighten.
4.
Turn on system and check for leaks.
SILICABUSTER MAINTENANCE AND REPLACEMENT:
Procedure:
1.
When the reading on the right-hand TDS meter (set to "OUT") displays
"001", it is time to replace the SilicaBuster™ DI cartridge.
2.
Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
3.
Tighten the cartridge housing by rotating it clockwise and hand tighten.
4.
Turn on system and check for leaks.
DM-1 METER Specifications:
Range
Resolution
Accuracy
Probe
Power Source
Battery life
For service or repair of these monitors, please send to:
SpectraPure
E-mail: sales@spectrapure.com Visit us on the web www.spectrapure.com
DI cartridge.
®
0 -1999 PPM
1PPM (1-999 PPM)
2%
0.25 inches
(2) 1.5V button batteries (A357 or equiv.)
Approx 1000 hours
HM DIGITAL, INC
5819 Uplander Way
Culver City, CA 90230
Inc .
®
Fax 480.894.6109 Fax us toll-free 1.877.527.7873
®
19

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MAXCAP RO/DI System and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents