MAXCAP DI MAINTENANCE AND REPLACEMENT:
Procedure:
1.
When the reading on the right-hand TDS meter (set to "IN") displays 75%
of the reading on the left-hand TDS meter (set to "OUT"), it is time to
replace the M
AX
Follow the directions on page 5 to rinse up the M
2.
cartridge
.
3.
Make sure the DI cartridge is installed in the correct direction as marked
on the cartridge shell and be sure that the top seal is securely attached to
the top of the cartridge.
4.
Tighten the cartridge housing by rotating it clockwise and hand tighten.
5.
Turn on system and check for leaks.
SILICA-BUSTER MAINTENANCE AND REPLACEMENT:
Procedure:
1.
When the reading on the right-hand TDS meter (set to "OUT") displays
"001", it is time to replace the SilicaBuster
2.
Follow the directions on page 5 to rinse up the SilicaBuster
3.
Make sure the DI cartridge is installed in the correct direction as marked
on the cartridge shell and be sure that the top seal is securely attached to
the top of the cartridge.
4.
Tighten the cartridge housing by rotating it clockwise and hand tighten.
5.
Turn on system and check for leaks.
DM-1 METER Specifications:
Range 0-1999 PPM
Accuracy
2%
Power Source
(2) 1.5V button batteries
For service or repair of these monitors, please send to:
SpectraPure®Inc .
E-mail: spectra@spectrapure.com Visit us on the web www.spectrapure.com
C
DI
cartridge
™
AP
Fax 480.894.6109 Fax us toll-free 1.877.527.7873
®
.
C
AX
AP
DI cartridge.
™
Resolution 1PPM
Probe
Battery life
HM DIGITAL, INC
5819 Uplander Way
Culver City, CA 90230
DI
™
DI cartridge.
™
(1-999 PPM)
¼''
Approx 1000 hours
19