Chapter 5 Technical Indicators; Appendix A : Technical Indicators - Tenma 72-8710 Operating Manual

Digital storage oscilloscope
Table of Contents

Advertisement

DSO Series User Ma nual
Chapter 5 Appendixes

Appendix A : Technical Indicators

Unless otherwise specified, all technical specifications apply to probes
with attenuation setting of 10X and the DSO Series Digital
Oscilloscope. To verify that your oscilloscope meets specifications, it
must first meet the following conditions :
n
The oscilloscope must have been operating continuously for
thirty minutes at the specified operating temperature.
n
If the operating temperature changes by more than 5°C, you
must perform the "Self Cal" operation, accessible through
the system functions menu. All specifications are guaranteed
unless noted "Typical".
Technical Indicators
Sampling
Sampling modes
72-8710
Real time sampling
Equivalent sampling
Bandwidth
100MHz
Sampling
Sampling, peak detect, average
When all channels have made N samplings
Average value
simultaneously, N is selectable from 2, 4, 8, 16,
32, 64, 128 to 256
Input
Input coupling
DC, AC, GND
1MΩ ± 2% in parallel with 20 ± 3pF
Input impedance
Probe attenuation factor
1 , 10×, 100×, 1000×
Maximum input voltage
400V (DC + AC Peak, 1MΩ input
impedance)
T i m e d e l a y b e t w e e n
150ps
channels (Typical)
72-8705
1GS/s
50GS/s
50MHz
×
81

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

72-8705

Table of Contents