Tool
Balance
Flexible hoses
Reference Manuals System
Binder I, II and III
The following chemicals and supplies must be supplied by the user:
Item
Water of suitable standard for use in the
system
Acetone solutions
Ethanol
Calibration solutions for pH
Conductivity solutions
ÄKTAprocess Site Preparation Guide 29-1396-55 AA
Specification
Certified and calibrated
The balance must be able to weigh enough solution
for the maximum flow rate for each system.
6 mm and 3/8": The balance must be able to
•
weigh 10 kg/2 minutes collect (maximum flow
rate: 180 l/h)
10 mm and 1/2": The balance must be able to
•
weigh 30 kg/2 minutes collect (maximum flow
rate: 600 l/h)
1" SS and PP: The balance must be able to weigh
•
80 kg/2 minutes collect (maximum flow rate:
2000 l/h)
Three (3) hoses (2 meters each) of the correct in-
•
ternal diameter
TC connections
•
System documentation is delivered with the system
and is required during installation.
Specification
For example, WFI/MQ purified
1.0% vol/vol
•
0.5% vol/vol
•
0.1% vol/vol
•
in sealed containers
20%
pH 4.00 (200 ml)
•
pH 7.00 (200 ml)
•
Approximately 2, 12 and 110 mS/cm
2 Site preparation overview
11
Need help?
Do you have a question about the ÄKTAprocess and is the answer not in the manual?
Questions and answers