Error Messages - Konica Minolta VIVID 9i/VI Instruction Manual

Non-contact 3d digitizer
Table of Contents

Advertisement

Error Messages

The following error messages appear if incorrect operation is performed or an abnormality occurs with the
VIVID 9i. If an error message appears, take the appropriate corrective actions as shown in the table below. If
the problem still does not disappear, contact a KONICA MINOLTA SENSING authorized service facility.
Error Message
(Alphabetically)
AF ERROR
LOW CONTRAST
PRESS ANY KEY
BACKUP ERROR
PRESS ANY KEY
BACKUP ERROR
PRESS ANY KEY AND SCAN
FIELD CALIBRATION SYS-
TEM
CAUTION
LASER SCANNER IS
OUT OF ORDER
PLEASE TURN OFF
CHANGE LENS
S/N ********
PRESS ANY KEY AND SCAN
FIELD CALIBRATION SYS-
TEM
LIGHT IMPROPER
SAME FILE NAME
SYSTEM ERROR
PLEASE TURN OFF
TERMINATE ERROR
PRESS ANY KEY
SET TERMINATE ENABLE
WB ERROR
LIGHT IMPROPER
PRESS ANY KEY
48
Meaning
Auto-focus failed due to low contrast
of the object. Press any key.
Failed to back up the data.
Press any key.
Failed to back up field calibration
values.
Press any key and scan field calibra-
tion system.
The laser beam is not emitted prop-
erly. Turn off the power immediately.
The lens is now being replaced.
Press any key and scan field calibra-
tion system.
Inadequacy amount of light
A file of the same file name already
exists.
A system error has occurred.
Turn OFF the power.
SCSI terminator error has occurred.
Press any key and set the terminator
to "ENABLE".
Failed to adjust white balance.
Press any key.
Action
Set the manual focus.
All the settings will be initialized, so set
them again.
Field calibration values have been reini-
tialized. Please try field calibration again.
Highly dangerous! Turn off the power
immediately and contact a KONICA
MINOLTA SENSING authorized service
facility.
Attach the lens hood properly.
Check the serial No. of the lens, and
replace the lens properly.
Illuminate the object brighter and scan it.
Specify another file name.
Turn OFF the power, then turn it ON
again. If the problem still does not disap-
pear, contact a KONICA MINOLTA
SENSING-authorized service facility.
Set the terminator to "ENABLE".
Adjust white balance in a bright place.
(White balance in effect before occur-
rence of this error has not been updated.)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 9i

Table of Contents