CI Maintenance

To Install the EI HES Radiator

Materials needed
• Gloves, clean, lint-free
• T10 Torx screwdriver (5182-3466)
• Tweezers (8710-2460)
Procedure
The analyzers, GC/MS interface, and other components in the analyzer chamber
WA R N I N G
operate at very high temperatures. Do not touch any part until you are sure it is cool.
Always wear clean gloves to prevent contamination when working in the analyzer
C A U T I O N
chamber.
Make sure you use an antistatic wrist strap and take other antistatic precautions
C A U T I O N
before touching analyzer components.
1
2
208
Large (8650-0030)
Small (8650-0029)
Place the radiator over the guide pins on the analyzer support and use a
T10 Torx screwdriver to secure it using the (2) M3 x 12 gold plated screws
(part number G7002-20110).
Use tweezers to disconnect the wires from the EI HES radiator. (See
Connect/Disconnect Wiring to the EI HES Radiator"
on page 201.)