To Install a Capillary Column in the GC/MS Interface Using a
Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self tightening
column nut explained in
Interface Using the Self-Tightening Column Nut"
Materials needed
• Interface tip seal (G3870-20542) (See
• Tip seal spring (G7005-20024) (Not used with SS or Inert EI source)
• Column cutter, ceramic (5181-8836) or diamond (5183-4620)
• Flashlight
• Magnifying loupe
• Gloves, clean
• Interface column nut (05988-20066)
• Ferrules
• Septum (may be old, used inlet septum)
• Safety glasses
• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
5977B Series MSD Operating Manual
page 44.) (Not used with EI SS or Inert source)
Large (8650-0030)
Small (8650-0029)
0.3 mm id, for 0.10 mm id columns (5062-3507)
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
0.5 mm id, for 0.32 mm id columns (5062-3506)
0.8 mm id, for 0.53 mm id columns (5062-3512)
"To Install a Capillary Column in the GC/MS
Figure 5
Installing 7890 GC Columns
on page 42.
on page 44 and
Figure 6
2
on
47