To Calibrate Column Flow Linear Velocity

The steps listed below apply to 7890 and 6890 series GCs only. Capillary
columns must be calibrated prior to use with the MSD.
Procedure
1 Set Data Acquisition for splitless manual injection, and set up a real time
2 Press [Prep Run] on the GC keypad.
3 Inject 1 µL of air into the GC inlet and press [Start Run].
4 Wait until a peak elutes at m/z 28. Note the retention time.
5 In the Instrument Control view, select GC Parameters from the Instrument menu.
6 Select the Configuration tab and then select the Columns tab.
7 Select your installed column from the table.
8 Click Calibrate to display the Calibrate Column dialog.
52
plot to monitor m/z 28.