To Connect/Disconnect Wiring to the EI HES Radiator

Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Pliers, long-nose (8710-1094)
• Tweezers (8710-2460)
Procedure
1 Use tweezers or needle nose pliers to connect/disconnect the green ground
2 Use tweezers or needle nose pliers to connect/disconnect the 2 purple
Entrance lens (1)*
Ion focus lens (2)*
Post extractor lens 2 (3)*
Post extractor lens 1 (4)*
Extractor lens (5)*
Ground wire
Heater wires
RTD wires
* The number in parenthesis is the
number engraved on the lens.
Figure 65
Wiring between the ceramic source board and the EI HES radiator
5977B Series MSD Operating Manual
wire and the 5 lens wires from the radiator. Do not bend the wires any more
than necessary. (See
source heater wires and the 2 gray RTD wires from the ceramic source board.
Figure
65.)
Ceramic source board
CI Maintenance
7