Appendix A Specifications; Defibrillator Specifications - ZOLL AED 3 Administrator's Manual

Defibrillator
Hide thumbs Also See for AED 3:
Table of Contents

Advertisement

Appendix A
Specifications

Defibrillator Specifications

This section describes product specifications for the ZOLL AED 3 defibrillator:
Device
Size (H x W x D)
Weight
Power
Device Classification
Design Standards
Environment
Operating Temperature
Storage Temperature
Humidity
Vibration
Shock
Altitude
Particle and Water Ingress
Drop
9650-000752-01 Rev. D
5.0" x 9.3" x 9.7" (12.7 cm x 23.6 cm x 24.7 cm)
5.5 lbs.; 2.5 kg
Battery pack
Internally powered per EN60601-1
Meets applicable requirements of EN 60601-1,
IEC 60601-1-11, EN 60601-2-4
32° to 122° F; 0° to 50° C
-22° to 158° F; -30° to 70° C
10 to 95% relative humidity, non-condensing
IEC 60068-2-64, Random, Spectrum A.4, Table A.8,
Cat. 3b; RTCA/DO-160G, Fixed Wing Aircraft,
Section 8.6, Test Cat. H, Aircraft Zone 1 and 2; EN 1789,
Sweep per EN 60068-2-6 Test Fc
IEC 60068-2-27; 100G
-1,250 to 15,000 ft.; -381 m to 4573 m
IP55
3.28 ft;1 m
ZOLL AED 3 Administrator's Guide