Specifications

Specifications
This section summarizes the specifications of the E Series End Tidal Carbon Dioxide (EtCO
Transducer Type
Principle of Operation
Warm Up Time
EtCO
Measurement Range
2
EtCO
Accuracy
2
(at 760 mmHg, ambient
temperature of 25°C)
EtCO
Resolution
2
EtCO
Stability
2
EtCO
Noise
2
EtCO
Rise Time (10-90%)
2
Respiration Rate Range
Respiration Rate Accuracy
Sample Flow Rate
Compensations
EtCO
Alarm Limits
2
Respiration Rate (RR) Alarm
Limits
Halogenated Agents
Airway Adapter Deadspace
Environmental (sensor or
module)
Electromagnetic Immunity
9650-1212-01 Rev. E
CAPNOSTAT 5® CO
Sensor
2
Mainstream
Non-Dispersive Infrared (NDIR) single beam optics, dual wavelength, no moving parts.
Full specifications within 2 minutes at an ambient temperature of 25º C. Capnogram in
15 seconds.
0 - 150 mmHg
0 - 20%
0 - 20 kPa
0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
101 - 150 mmHg, ±10% of reading
(At respiration rates > 80 breaths per minute, all ranges are ±12% of actual.)
0.1 mmHg 0 - 69 mmHg
0.25 mmHg 70 - 150 mmHg
Short-Term Drift: Drift over four hours < 0.8 mmHg.
Long-Term Drift: Accuracy specification will be maintained over a 120 hour period after
zeroing.
RMS noise of the sensor < 0.25 mmHg at
7.5% CO
.
2
< 60 ms (Adult/pediatric adapters)
< 60 ms (Infant/pediatric adapters)
2 - 150 breaths per minute
±1 breath per minute
N/A
Barometric pressure 400 - 850 mmHg (automatic).
Operator selectable O
/N
O compensation.
2
2
User selectable, High 5 - 100 mmHg, Low 0 - 95 mmHg, OFF.
User selectable, High 5 - 150 respirations per minute, Low 0 - 100 respirations per
minute, OFF.
Specification allows for halogenated anesthetic agents that may be present at normal
clinical levels. The presence of Desflurane in the exhaled breath beyond normal values
(5-6%) may positively bias carbon dioxide values by up to an additional 2-3 mmHg.
Adult < 5 cc
Infant < 1.0 cc
Operating Temperature: 0º C to 45º C (0º C to 40º C for LoFlo Module)
Storage and Shipping Temperature: -40º C to 70º C
The E Series unit may not perform to specifications when stored at the upper or lower
extreme limits of storage temperature and immediately put into use.
AAMI DF-80:2003, IEC 60601-1-2, 10 V/m
) option.
2
LoFlo™
Sidestream
RMS noise of the sensor < 0.25 mmHg at
5% CO
.
2
< 3 seconds (includes transport and rise
time)
50 ml/min ±10 ml/min
Adult < 7.3 cc Maximum
Pediatric/Infant < 1.0 cc