This section summarizes the specifications of the M Series End-Tidal Carbon Dioxide (EtCO
Transducer Type
Principle of Operation
Warm Up Time
EtCO
Measurement Range
2
EtCO
Accuracy
2
(at 760 mmHg, ambient
temperature of 25°C)
EtCO
Resolution
2
EtCO
Stability
2
EtCO
Noise
2
EtCO
Rise Time (10-90%)
2
Respiration Rate Range
Respiration Rate Accuracy
Sample Flow Rate
Compensations
EtCO
Alarm Limits
2
Respiration Rate (RR) Alarm
Limits
Halogenated Agents
Airway Adapter Deadspace
Environmental
Electromagnetic Immunity
CAPNOSTAT 3
Mainstream
Non-Dispersive Infrared (NDIR) single beam optics, dual wavelength, no moving
parts.
Full specifications within 60 seconds. Capnogram in 15 seconds.
0 - 100 mmHg
0 - 13%
0 - 12.5 kPa
0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
0.5 mmHg
Short Term Drift: Drift over four hours < 0.8 mmHg.
Long Term Drift: Accuracy specification will be maintained over a 120 hour period
after zeroing.
RMS noise of the sensor < 0.25 mmHg at 7.5% CO
< 60 ms (Adult/pediatric adapters)
< 50 ms (Neonatal adapters)
0 - 150 breaths per minute
±1 breath per minute
N/A
Barometric pressure 550 - 780 mmHg (automatic).
Operator selectable O
/N
O compensation.
2
2
User selectable, High 5 - 100 mmHg, Low 0 - 95 mmHg, OFF.
User selectable, High 5 - 150 respirations per minute, Low 0 - 100 respirations per
minute, OFF.
Specification allows for halogenated anesthetic agents that may be present at
normal clinical levels. The presence of Desflurane in the exhaled breath beyond
normal values (5-6%) may positively bias carbon dioxide values by up to an
additional 2-3 mmHg.
Adult < 5 cc
Infant < 1.0 cc
Operating Temperature:
Storage and Shipping Temperature: -10º C to 55º C
The M Series unit may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.
AAMI DF-39, IEC 60601-1-2, 20 V/m
EtCO
- 21
2
End Tidal Carbon Dioxide (EtCO
) option.
2
LoFlo
Sidestream
0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
±12% for respiration rate above
80 BrPM
.
2
< 200 ms
50 ml/min
Adult 7.3 cc maximum
Pediatric/Infant < 1 cc
10º C to 40º C
)