Instrument; Trigger - LeCroy LogicStudio Getting Started Manual

Logic analyzer
Table of Contents

Advertisement

Getting Started Manual 

Instrument

The Instrument setup button shows a configuration screen where you can configure how the digital instrument 
samples data ‐ including the half/full channel mode selection and setting threshold levels. Changes made on the 
Instrument setup screen take effect on the next acquisition. 
 
Note: If you select a different channel mode, the information shown on the left of this setup display is updated 
to match acquisition details for your choice. 

Trigger

The Trigger setup button opens a pop‐up used for configuring trigger conditions for the digital device. The 
trigger pop‐up has four tabs for different trigger categories. 
Note: It's important to note that the trigger logic always runs at a rate of 500 MHz under all modes, regardless 
of the selected Sample Rate. This allows you to continue triggering on fast events even when lowering the 
Sample Rate to obtain a longer timeframe of data. However, this may result in trigger events that cannot be 
seen in the slower sampled acquisition data when presented on the waveform display area. 
S
L
T
 
IMPLE 
OGIC 
AB
Use this tab to setup simple, logic‐based triggers. Modes include Edge, Level, Pattern, and Immediate triggers. 
A
L
T
 
DVANCED 
OGIC 
AB
Use this tab to set up more complex logic‐based triggers. You can set the trigger mode from Edge(s), Width, 
Level(s), Qualified Level(s), Pattern, and Qualified Pattern choices. A different trigger mode can also be selected 
for each of two trigger blocks (A and B) and then they can be combined in different logical ways including: A 
only, B only, A and B, A or B, A then B, and B then A. 
 
24