Control Drawing Of Sc202S Ff/Pb Specification - YOKOGAWA SC202G User Manual

2-wire conductivity or resistivity transmitter
Hide thumbs Also See for SC202G:
Table of Contents

Advertisement

2-12. Control Drawing of SC202S FF/PB Specification (FM Intrinsically safe FISCO)
FM Approved
FISCO barrier
Voc (Vt) 17,5 V
+
Ioc (It) 380 mA
Poc (Pt) 5,32 W
-
Unclassified Location
Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or
energy over 20 J, or are FM Approvals entity approved and meet connection
requirements.
Electrical data of the SC202S-F & SC202S-P:
Supply circuit: Vmax = 17.5 V; Imax = 380 mA; Pi = 5.32 W; Ci = 220 pF; Li = 0 H.
-
Sensor input circuit: Vt = 14.4 V; It = 10 mA; Ca = 59.36 nF; La = 340 mH
-
Any FM Approved FISCO barrier may be used that meets the following requirements:
Voc or Vt
17.5 V; Ioc or It
When installing this equipment, follow the manufacturer's installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 "Installation of
Intrinsically Safe Systems for Hazardous (Classified) Locations" and the National
Electrical Code (ANSI/NFPA 70).
Associated apparatus connected to the FISCO barrier must not use or generate more
than 250 Vrms or Vdc.
Resistance between FISCO Intrinsically Safe Ground and earth ground must be less than
1.0 Ohm.
In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
The FISCO concept allows the interconnection of several I.S. apparatus not specifically
examined in such combination. The criterion for such interconnection is that the voltage
(Vmax), the current (Imax) and the power (Pi) which I.S. apparatus can receive and remain
intrinsically safe, considering faults, must be equal to or greater that the voltage (Voc, Vt),
the current (Ioc, It) and the power (Poc, Pt) which can be providede by the FM approved
FISCO barrier. In addition, the maximum unprotected residual capacitance (Ci) and
inductance (Li) of each apparatus (other than the terminator) connected to the Fieldbus
must be less than or equal to 5nF and 10 μH respectively.
FM Class I, DIV. 1, Group ABCD
T4 for ambient temp.
SC202S-F
or SC202S-P
FM Approved
Terminator
R = 90..100Ω
C = 0..2,2 μF
380 mA; Poc or Pt
55 C
-
+
-
+
Transmitter
Division 1
Classified Location
5.32 W
Specifications 2-15
Sensor
Connections
Max. cablelength: 60 mtr.
Cable dia. : 3...12 mm.
Sensor
Connections
FM Approved
Terminator
R = 90..100Ω
C = 0..2,2 μF
-
+
Transmitter
IM 12D08B02-01E