Sc202S 2-Wire Conductivity Transmitter - YOKOGAWA SC202G User Manual

2-wire conductivity or resistivity transmitter
Hide thumbs Also See for SC202G:
Table of Contents

Advertisement

13-2. SC202S 2-Wire Conductivity Transmitter

Quality
Inspection
Standards
1.
Scope
This inspection standard applies to the SC202
2.
Inspection Items
2.1 Insulation resistance test
* 2.2 Dielectric strength test
2.3 Current output test
2.4 Temperature indication check
2.5 Resistance (conductivity) check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3.
Inspection Methods, Standards and Conditions
Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test,
follow the instructions in Section 3.1 and for the connections for the dielectric strength test,
follow the instructions in Section 3.2.
Performance tests should be done in the inspection mode where the tests from Section 3.3
through Section 3.5 take place in sequence and cannot be retraced. If the reconfirmation of
a test is needed, turn off the power to the transmitter, turn on the power again, and enter the
inspection mode to restart the tests.
Set the equipment as follows.
Decade resistance box 1 (temperature): 960.9
Decade resistance box 2 (conductivity):
DC source:
3.1
Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 MΩ or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 16) and the earth
terminal (G). The insulation resistance must be 100 MΩ or greater.
3.2
Dielectric strength test
Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz
or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must
withstand this voltage. (The sensed current should be 10 mA.)
(1) Between the power supply terminals shorted together (+ and –) and the earth terminal (G)
(2) Between the input terminals shorted together (11 to 16) and the earth terminal (G)
(3) Between the input terminals shorted together (11 to 16) and the power supply terminals
shorted together (+ and –)
3.3
Current Output Test
<Test>
(1) Enter Service Code 87 and then password 070.
(2) When the message display shows "*HIF", press the [YES] key.
SC202S
2-Wire Conductivity Transmitter
2-Wire Conductivity Transmitter.
10
24 VDC
Appendix 3-5
QIS 12D08B02-21E
1st Edition: Mar. 2007(YK)
IM 12D08B02-01E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents