Calibration Procedure - YOKOGAWA SC202G User Manual

2-wire conductivity or resistivity transmitter
Hide thumbs Also See for SC202G:
Table of Contents

Advertisement

6-2 Calibration

6-2. Calibration procedure

YES
NO
MODE
YES
NO
YES
YES
NO
Put the sensor in standard
solution. Press YES.
After the indication is stable, set the value
using the >, , ENT key.
The cell constant is automatically updated after the calibration and the new
value can be read on the display as described in section 4.5.
The calculation is as follows: Cell constant in /cm= (Conductivity of calibration solution in mS/cm) x
(Cell resistance in kOhm)
Comparing this calibrated cell constant with the initial nominal cell constant in service code 03
gives a good indication of the stability of the sensor. If the calibrated cell constant differs more than
20% from the nominal cell constant error E3 is displayed.
IM 12D08B02-01E
MODE
MEASURE
CAL
DISPLAY
HOLD
MODE
ENT
Select the flashing digit with the > key.
Increase its value by pressing the key
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
ENT
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
ENT
YES
ENT
NO